15W LH15-20BxxMU and 25W LH25-20BxxMU series targeting medical application feature reinforced insulation design, 4000VAC high isolation and low leakage current (<100µA) together with output short-circuit, over-current and over-voltage protections to ensure the safety of patients and the system. Both series meet medical design standard EN60601-1 and ANSI/AAMI ES60601-1 (3rd edition, 2×MOPP).
The series also offer high efficiency up to 89% and low standby power consumption less than 0.1W to save energy effectively. Moreover, they provide operating temperature of -40℃ to +70℃ and high operating elevation up to 5000m, meeting requirements of harsh environment.
Applications :
Medical diagnosic/auxiliary equipment: injection pump, multi-functional monitoring device and incubator,etc.
Features:
Input voltage range: 85-264VAC/100-370VDC
Standby power consumption<0.1W
Operating temperature: -40℃ to +70℃
Isolation: 4000VAC
High efficiency up to 89%
Reinforced insulation design
Output short-circuit, over-current and over-voltage protections
Meet EN60601-1, ANSI/AAMI ES60601-1 (2xMOPP)
Optional packages: PCB mounting, chassis mounting, DIN-Rail mounting
Leakage current: <100µA (RMS typ./240VAC/63Hz)
Operating elevation up to 5000m